OTDR WizardSign in

Reference · 4 min read

Cumulative loss vs total loss

Updated

Total loss is the end-to-end loss the OTDR measured and stored in the file: the drop in power between the start and the end of the fibre, taken from the trace itself. Cumulative loss is a running total worked out from the event table: each fibre section’s length times its attenuation, plus each event’s loss, added up from the start. The two often agree within a few hundredths of a dB, but they are made in different ways, so they can differ. For acceptance and reports, quote the total loss.

Two numbers, two methods

AspectTotal lossCumulative loss
Where you see itThe Total loss tile and the report resultsThe last column of the event table
Made byThe OTDR, when it analysed the traceThe viewer, from the event table
MethodPower level at the start minus power level at the end of the fibreSum of section losses and event losses
Stored in the fileYes, in the KeyEvents block, as end-to-end lossNo, calculated
Includes small events below the detection thresholdYes, they are part of the measured dropNo, only the listed events

How the cumulative loss adds up

For every event, from the launch point:

Cumulative loss = previous cumulative loss + section length × section attenuation + event loss

When the two agree

This 1310 nm trace from an EXFO MaxTester 730C adds up to almost exactly what the instrument measured:

#TypeSection km× Att. dB/km= Fibre dB+ Event dBCumulative loss dB
1Reflective———0.0000.000
2Reflective0.15030.6870.1030.6520.755
3Reflective3.58890.3221.1560.0001.911
4Reflective0.1733—0.0000.0001.911
5Reflective3.4150—0.0000.0001.911
6Fibre end0.1743—0.000—1.911
Cumulative loss at the fibre end (calculated)
1.91 dB
Total loss stored by the OTDR (measured)
1.91 dB
Difference
0.00 dB
How the cumulative loss adds up for EXFO MaxTester 730C, 1310 nm.

The cumulative loss at the end is 1.91 dB and the OTDR measured 1.91 dB. The two match because the instrument’s own section attenuations and event losses describe the whole fibre.

When they do not

This 1550 nm trace from a Noyes OFL280 does not:

#TypeSection km× Att. dB/km= Fibre dB+ Event dBCumulative loss dB
1Reflective———-0.215-0.215
2Non-reflective0.0109—0.0000.3740.159
3Fibre end3.72360.1850.689—0.848
Cumulative loss at the fibre end (calculated)
0.85 dB
Total loss stored by the OTDR (measured)
0.58 dB
Difference
0.27 dB
How the cumulative loss adds up for Noyes OFL280, 1550 nm.

The cumulative loss reaches 0.85 dB, while the OTDR stored 0.58 dB, a difference of 0.27 dB. The table shows why the two methods drift apart here: the first event has a negative loss of -0.21 dB, the second sits only 0.0109 km from it with 0.37 dB, and almost all of the rest comes from one long section multiplied by one attenuation value. The instrument instead measured the drop in power across the fibre, between its own start and end markers.

Why they differ

  • Hidden events. An OTDR only lists events above its loss threshold, often 0.05 to 0.2 dB. Smaller splices and gentle bends are not in the table, but their loss is in the measured total.
  • Near-end events. Events in or just after the launch connector’s dead zone are hard to measure. A negative or uncertain loss there shifts the whole running total.
  • Missing attenuation. If a section has no stored attenuation and none can be measured from the trace, it adds nothing to the cumulative loss, while its real loss is still in the measured total.
  • Markers. The instrument’s total runs between its start and end markers. Where those sit, before or after the first connector, and at or just before the fibre end, changes the result by the loss of those connectors.
  • Rounding. Each event loss is stored to 0.001 dB and each attenuation to 0.001 dB/km. Over many events and long sections, small rounding adds up.

Which one to report

  • For acceptance and reports, use the total loss. It is what the instrument measured, it is what its own report shows, and it is what a technician sees on the OTDR screen.
  • Use the cumulative loss to see where the loss builds up, for example which part of a route uses most of the budget, or how much loss is left after a given splice closure.
  • For a loss budget check, OTDR Wizard compares the cumulative loss at the fibre end with your Max. total cumulative loss limit; see pass/fail limits. If your specification is written against measured end-to-end loss, compare it with the Total loss tile as well.

For the most accurate link loss, many specifications also ask for a light source and power meter test (an optical loss test set) in addition to the OTDR trace: it measures the loss of the whole link, connectors included, directly.